GE LOGIQ S8 system maintenance

2024-07-01
Latest company news about GE LOGIQ S8 system maintenance

GE LOGIQ S8 system maintenance

 

1. Failure phenomenon:

2D image and blood flow of the machine is abnormal, the blood flow knob is not working good.

 

2. Failure analysis:

According to the above failure phenomenon, it can be judged that the image channel of the machine is abnormal. Therefore, it is suspected that there is a problem with the BF channel board.

It was also found that the knob& of blood flow section was not working good, so the control panel was disassembled and the circuit board of the blood flow button needs to be repaired.

 

3. Result:
After replace the BF channel board and repair the circuit board of& the blood flow button, everything back to normal.

products
NEWS DETAILS
GE LOGIQ S8 system maintenance
2024-07-01
Latest company news about GE LOGIQ S8 system maintenance

GE LOGIQ S8 system maintenance

 

1. Failure phenomenon:

2D image and blood flow of the machine is abnormal, the blood flow knob is not working good.

 

2. Failure analysis:

According to the above failure phenomenon, it can be judged that the image channel of the machine is abnormal. Therefore, it is suspected that there is a problem with the BF channel board.

It was also found that the knob& of blood flow section was not working good, so the control panel was disassembled and the circuit board of the blood flow button needs to be repaired.

 

3. Result:
After replace the BF channel board and repair the circuit board of& the blood flow button, everything back to normal.